Evaluación del desempeño de la impedancia en modelos estructurales de electrodos EEG de tecnología MNA
Date
2021-02-05Authors
Sulez Gómez, Felipe AntonioPublisher
Pontificia Universidad Javeriana
Faculty
Facultad de Ingeniería
Program
Maestría en Ingeniería Electrónica
Obtained title
Magíster en Ingeniería Electrónica
Type
Tesis/Trabajo de grado - Monografía - Maestría
COAR
Tesis de maestríaShare this record
Citación
Metadata
Show full item record
PDF documents
English Title
Evaluation of impedance performance in structural models of EEG electrodes of MNA technologyResumen
El desempeño de la impedancia como parámetro eléctrico en modelos estructurales de electrodos de Electroencefalografía (EEG) es muy importante para adquirir señales biológicas de calidad. Las curvas de impedancia se han medido en simulación en la superficie del cuero cabelludo entre el electrodo y la piel. El comportamiento de la curva de impedancia es de tipo asintótico decreciente en un rango de frecuencia de 0 Hz a 1.5 MHz. Para determinar las variaciones de impedancia, se construyeron 4 modelos de electrodos de matriz de microagujas (MNA), utilizando un método de elementos finitos (FEM). Por lo anterior, los resultados reportados en la literatura científica recomiendan el uso de esta tecnología en aplicaciones biomédicas, debido a la baja impedancia mantenida por los electrodos MNA en el tiempo con relación a los electrodos húmedos estándar. En este sentido, se lleva a cabo la comparación de curvas de impedancia entre el modelo de electrodo FEM y el modelo húmedo estándar. Como resultado, la impedancia entre la piel y el cuero cabelludo es de aproximadamente 7.5 kohms. Se desea que con el uso de la tecnología MNA la media de la impedancia se haya ubicado en el rango de impedancia permisible. Utilizando Ansys Electronics ™, se diseñaron y evaluaron modelos de electrodos MNA, y sus resultados se contrastaron con los principales modelos de la literatura. Además, se identificó que la variación de los parámetros estructurales evaluados conduce a una disminución de la impedancia y permitirá definir una nueva topología de electrodo, que reducirá el valor de la impedancia de acuerdo con los rangos permisibles para la adquisición de EEG y su caracterización recomendada para implementar en aplicaciones de EEG de grabación biomédica.
Abstract
The performance of impedance as an electrical parameter in structural models of electrodes of Electroencephalography (EEG) is very important to acquire a good quality of biological signals. The curves of impedance have been measured in simulation on the surface of the scalp between electrode and skin. The behavior of the impedance curve is a decreasing asymptotic type in a frequency range of 0 Hz to 1.5 MHz. To determine impedance variations, 4 Microneedle Array (MNA) electrode models were constructed, using a Finite Element Method (FEM).. The curves of impedance have been measured in simulation on the surface of the scalp between electrode and skin. The behavior of the impedance curve is a decreasing asymptotic type in a frequency range of 0 Hz to 1.5 MHz. To determine impedance variations, 4 Microneedle Array (MNA) electrode models were constructed, using a Finite Element Method (FEM). Due to the above, results reported in the scientific literature recommend the use of this technology in biomedical applications, due to low impedance maintained by MNA electrodes over time relative to standard wet electrodes. In this sense, is carried out the comparison of impedance curves between the FEM electrode model and the standard wet model. As a result, the impedance between the skin and scalp is approximately 7.5 kohms. It is desired that with the use of the MNA technology the mean of the impedance has been located in the impedance range permissible. Using Ansys Electronics™, MNA electrode models were designed and evaluated, and their results were contrasted with the main models of the literature. Further, it was identified that the variation of the evaluated structural parameters leads to a decreased impedance and will allow defining a new electrode topology, which will reduce the value of the impedance according to the allowable ranges, for EEG acquisition and its recommended characterization to implement in biomedical recording EEG applications
Keywords
Interfaz electrodo pielModelo electrodo estándar
Modelo electrodo seco
Electroencefalografía
Método de elementos finitos
Keywords
Skin electrode interfaceDry electrodes
Wet electrodes
Conductive gel
Finite element method
Themes
Maestría en ingeniería electrónica - Tesis y disertaciones académicasElectroencefalografía
Electrodos
Impedancia de transferencia